An in vitro model test system for estimation of the blood compatibility of hemodialysis membranes and tubing is described. The model test system consists of a modified hemodialysis unit and blood pump through which fresh citrated human blood is circulated. The effects of the use of different pump and tubing types upon hematologic and blood coagulation parameters are described. Preexposure of test surfaces to albumin appeared to enhance blood compatibility characteristics of the model test system, whereas preexposure to a high density lipoprotein preparation or a proteinpolysaccharide preparation was without appreciable benefit. Use of blood from subjects receiving aspirin resulted in enhanced blood compatibility in the test system as did use of heparin. Use of Warfarin or dextran did not appear to enhance blood compatibility of test surfaces under the conditions of this test system. Dialysis membranes and tubing which formed parts of the test system were examined by scanning and transmission electron microscopy in control tests and in tests for effects of proteins and antithrombotic agents.