[Non-invasive blood pressure measuring device].

N I Kaliadin, V A Lemenkov, A V Korobeĭnikov, S M Perevozchikov, V G Vlasov
Author Information

Abstract

Algorithms have been developed for the Russian non-invasive blood pressure measuring monitor MPK-01 that applies a frequency excess pressure transducer. The monitor uses oscillometry. A preliminary estimate of the accuracy of blood pressure measurement for its conformity to the ANSI/AAMI (USA) and BHS (UK) standards has indicated the legitimacy of the algorithm proposed for processing if there are hardware restrictions inherent in this monitor.

MeSH Term

Algorithms
Blood Pressure
Blood Pressure Determination
Blood Pressure Monitors
Humans
Oscillometry
Reference Values
Reproducibility of Results
Signal Processing, Computer-Assisted
Transducers, Pressure

Word Cloud

Created with Highcharts 10.0.0pressurebloodmonitormeasuringAlgorithmsdevelopedRussiannon-invasiveMPK-01appliesfrequencyexcesstransducerusesoscillometrypreliminaryestimateaccuracymeasurementconformityANSI/AAMIUSABHSUKstandardsindicatedlegitimacyalgorithmproposedprocessinghardwarerestrictionsinherent[Non-invasivedevice]