Accuracy of stent measurements using ECG-gated greyscale intravascular ultrasound images: a validation study.

Ryan K Kaple, Kenichi Tsujita, Akiko Maehara, Gary S Mintz
Author Information
  1. Ryan K Kaple: The Cardiovascular Research Foundation, Columbia University Medical Center, New York City, NY, USA.

Abstract

Greyscale intravascular ultrasound (IVUS) is an accurate tool for measuring stent dimensions and residual disease at the stent edge. Electrocardiographically (ECG)-gated IVUS is used in evolving second-generation IVUS systems, but this modality provides fewer greyscale cross-sectional images, and its accuracy to measure stent dimensions has not been assessed. This study was designed to validate the use of ECG-gated greyscale IVUS in measuring minimum stent area (MSA), stent length and reference dimensions compared to standard greyscale IVUS data. IVUS imaging was performed after drug-eluting stent implantation in 53 target lesions in 48 patients with acute coronary syndrome. The IVUS catheter was mechanically withdrawn at 0.5 mm/s, standard greyscale images were collected at 10 frames/s, and ECG-gated greyscale images were constructed from R-wave gated images. The MSA measured 6.20 +/- 1.75 mm(2)vs. 5.98 +/- 1.55 mm(2) on standard greyscale and ECG-gated greyscale IVUS, respectively (R(2) = 0.91, p = 0.005). The MSA position (R(2) = 0.66, p = 0.179) and stent length (R(2) = 0.99, p = 0.435) measurements were similar between modalities. Proximal reference vessel area was larger by ECG-gated IVUS, but proximal reference lumen and distal reference measurements were similar. Bland-Altman plots demonstrated good agreement between modalities. In conclusion, ECG-gated greyscale IVUS provides accurate and reliable measurements of stent length, area and reference segment plaque burden after stent implantation and is not inferior to standard greyscale IVUS.

MeSH Term

Acute Coronary Syndrome
Adult
Aged
Blood Vessel Prosthesis Implantation
Drug-Eluting Stents
Electrocardiography
Female
Humans
Image Processing, Computer-Assisted
Male
Middle Aged
Regression Analysis
Sensitivity and Specificity
Ultrasonography, Interventional

Word Cloud

Created with Highcharts 10.0.0IVUSstentgreyscale0ECG-gated=reference2imagesstandardmeasurementsdimensionsareaMSAlengthRpintravascularultrasoundaccuratemeasuringprovidesstudyimplantation5+/-1mmsimilarmodalitiesGreyscaletoolresidualdiseaseedgeElectrocardiographicallyECG-gatedusedevolvingsecond-generationsystemsmodalityfewercross-sectionalaccuracymeasureassesseddesignedvalidateuseminimumcompareddataimagingperformeddrug-eluting53targetlesions48patientsacutecoronarysyndromecathetermechanicallywithdrawnmm/scollected10frames/sconstructedR-wavegatedmeasured62075vs9855respectively91005position6617999435ProximalvessellargerproximallumendistalBland-AltmanplotsdemonstratedgoodagreementconclusionreliablesegmentplaqueburdeninferiorAccuracyusingimages:validation

Similar Articles

Cited By