Evaluation of Preoperative Implant Placement in Total Shoulder Arthroplasty.

Matthew A Hamilton, Sandrine Polakovic, Paul Saadi, Richard B Jones, Ira M Parsons, Emilie V Cheung
Author Information

Abstract

INTRODUCTION: New technology to assist with glenoid placement in shoulder arthroplasty has evolved to include preoperative planning tools and intraoperative guides. These tools provide surgeons with a more complete understanding of glenoid anatomy prior to surgery. However, there have been no studies identifying the information that most influences surgical decision making. Further, there have been few studies that quantify intraoperative identification of scapular landmarks required to execute a preoperative plan. The purpose of this study is to examine the variables that are considered when making a preoperative plan in shoulder arthroplasty.
METHODS: The first part of this study was a cadaveric lab in which three surgeons identified the neutral axis in surgical simulation. The second part of the study utilized a preliminary software tool in which surgeons were able to place glenoid implants in a set of CT reconstructions utilizing standard pegged glenoid components. In the third part of the study, surgeons utilized a novel planning software that included the ability to view the 3D reconstructed glenoid in all planes simultaneously and place either standard or augmented glenoid implants. The results of these three studies were compared.
RESULTS: The center of the glenoid identified in the cadaver lab was 1.69 mm ± 1.58 mm anterior and 1.99 mm ± 2.49 mm superior to center. The identified neutral axis was tilted 14.2° ± 9.2° superior to the Friedman axis with 11.8° ± 7.9° of retroversion relative to that axis. Using the novel preoperative planning tool, the surgeons placed implants less than 0.5 mm from the center of the glenoid (AP = -0.07 mm ± 0.42 mm, SI = 0.44 mm ± 0.82 mm) with an average retroversion of less than 1° (-0.96° ± 3.04°).
CONCLUSION: There was a discernible difference between the neutral axis identified in the cadaveric simulation (aver age of 14.2° superior and 11.8° retroverted) and the implant orientation planned using preoperative software (average of 3.26° superior and 0.96° retroverted). Based on the variability of position and orientation seen cadaverically, it is concluded that additional intraoperative guidance is needed alongside a preoperative plan in order to execute ideal placement of the glenoid component.

MeSH Term

Arthroplasty, Replacement
Cadaver
Computer Simulation
Glenoid Cavity
Humans
Imaging, Three-Dimensional
Joint Prosthesis
Prosthesis Design
Radiographic Image Interpretation, Computer-Assisted
Shoulder Joint
Software
Surgery, Computer-Assisted
Tomography, X-Ray Computed

Word Cloud

Created with Highcharts 10.0.0glenoidmm±preoperativesurgeonsaxis0studyidentifiedsuperiorplanningintraoperativestudiesplanpartneutralsoftwareimplantscenter1placementshoulderarthroplastytoolssurgicalmakingexecutecadavericlabthreesimulationutilizedtoolplacestandardnovel1411retroversionless=-0average96°3retrovertedorientationINTRODUCTION:NewtechnologyassistevolvedincludeguidesprovidecompleteunderstandinganatomypriorsurgeryHoweveridentifyinginformationinfluencesdecisionquantifyidentificationscapularlandmarksrequiredpurposeexaminevariablesconsideredMETHODS:firstsecondpreliminaryablesetCTreconstructionsutilizingpeggedcomponentsthirdincludedabilityview3DreconstructedplanessimultaneouslyeitheraugmentedresultscomparedRESULTS:cadaver6958anterior99249tilted9Friedman7relativeUsingplaced5AP0742SI448204°CONCLUSION:discernibledifferenceaverageimplantplannedusing26°BasedvariabilitypositionseencadavericallyconcludedadditionalguidanceneededalongsideorderidealcomponentEvaluationPreoperativeImplantPlacementTotalShoulderArthroplasty

Similar Articles

Cited By (1)