Abstract

Space charge effects play important roles in ion trap operations, which typically limit the ion trapping capacity, dynamic range, mass accuracy, and resolving power of a quadrupole ion trap. In this study, a rhombic ion excitation and ejection method was proposed to minimize space charge effects in a linear ion trap. Instead of applying a single dipolar AC excitation signal, two dipolar AC excitation signals with the same frequency and amplitude but 90° phase difference were applied in the x- and y-directions of the linear ion trap, respectively. As a result, mass selective excited ions would circle around the ion cloud located at the center of the ion trap, rather than go through the ion cloud. In this work, excited ions were then axially ejected and detected, but this rhombic ion excitation method could also be applied to linear ion traps with ion radial ejection capabilities. Experiments show that space charge induced mass resolution degradation and mass shift could be alleviated with this method. For the experimental conditions in this work, space charge induced mass shift could be decreased by ~50%, and the mass resolving power could be improved by ~2 times at the same time. Graphical Abstract ᅟ.

Keywords

References

  1. Anal Chem. 2014 Oct 21;86(20):10164-70 [PMID: 25252102]
  2. J Mass Spectrom. 2013 Aug;48(8):937-44 [PMID: 23893641]
  3. J Am Soc Mass Spectrom. 2010 Feb;21(2):203-8 [PMID: 19896390]
  4. J Mass Spectrom. 2005 Apr;40(4):430-43 [PMID: 15838939]
  5. JALA Charlottesv Va. 2010 Dec;15(6):433-439 [PMID: 21278840]
  6. Analyst. 2015 May 21;140(10):3406-14 [PMID: 25860539]
  7. Anal Chem. 1984 Dec;56(14 ):2744-8 [PMID: 6524653]
  8. J Mass Spectrom. 2004 May;39(5):471-84 [PMID: 15170743]
  9. Anal Chem. 2000 Mar 1;72(5):970-8 [PMID: 10739200]
  10. J Am Soc Mass Spectrom. 2002 Jun;13(6):659-69 [PMID: 12056566]
  11. J Am Soc Mass Spectrom. 2014 Mar;25(3):498-508 [PMID: 24385397]
  12. Rapid Commun Mass Spectrom. 2011 Dec 15;25(23):3509-20 [PMID: 22095499]
  13. Mass Spectrom Rev. 2005 Jan-Feb;24(1):1-29 [PMID: 15389865]
  14. Anal Chem. 2006 Apr 1;78(7):2113-20 [PMID: 16579588]
  15. J Am Soc Mass Spectrom. 2003 Oct;14(10):1130-47 [PMID: 14530094]
  16. J Am Soc Mass Spectrom. 2012 Oct;23(10):1799-807 [PMID: 22875333]
  17. Anal Chem. 2015 Feb 17;87(4):2236-41 [PMID: 25597704]
  18. Rapid Commun Mass Spectrom. 2007;21(22):3527-46 [PMID: 17944004]
  19. J Am Soc Mass Spectrom. 1994 Feb;5(2):64-71 [PMID: 24222517]

Word Cloud

Created with Highcharts 10.0.0iontrapmasschargeexcitationSpaceeffectsmethodspacelinearIonresolvingpowerrhombicejectiondipolarACappliedexcitedionscloudworkinducedshiftLinearRhombicplayimportantrolesoperationstypicallylimittrappingcapacitydynamicrangeaccuracyquadrupolestudyproposedminimizeInsteadapplyingsinglesignaltwosignalsfrequencyamplitude90°phasedifferencex-y-directionsrespectivelyresultselectivecirclearoundlocatedcenterrathergoaxiallyejecteddetectedalsotrapsradialcapabilitiesExperimentsshowresolutiondegradationalleviatedexperimentalconditionsdecreased~50%improved~2timestimeGraphicalAbstractReducingChargeEffectsTrapExcitationEjectiontrajectorysimulation

Similar Articles

Cited By