Abstract

BACKGROUND: Changes in the microcirculation may be used as a surrogate outcome in studies on cardiovascular disease. We assessed the reliability characteristics of the sublingual microcirculation parameters Vascular Density (VD), Red Blood Cell Filling (RBCF), and Perfused Boundary Region (PBR) as obtained by sidestream darkfield imaging.
METHODS: For each of the three parameters, the variance components of measurement, the Intraclass Correlation Coefficient (ICC), the Standard Error of Measurement, and the limits of agreement were estimated for the intra-rater setting (N = 50) and the inter-rater setting (N = 48). Subsequently, as a proof of concept, the reliability measures were used for a power analysis to design studies to evaluate the effect of acute stimuli-i.e. having a meal (N = 50) and cigarette smoking (N = 21) on the three parameters.
RESULTS: Reproducibility was poor for all three parameters. The intra-rater ICC for 2 measurements was 0.28 (95% CI: 0.04, 0.53) for the VD, 0.51 (95% CI: 0.27, 0.69) for the RBCF, and 0.33 (95% CI: 0.08-0.56) for the PBR. The standard errors of measurement and the limits of agreement for all three parameters were larger than most statistically significant intra-individual or inter-individual differences reported in previous studies. The proofs of concept showed that sample sizes in excess of 600 subjects are necessary to reach statistical significance for the observed effects of having a meal or smoking on VD and PBR.
CONCLUSIONS: The reliability of the three sublingual microcirculation parameters in their current form appears to be low and a large sample size is advisable for their use in conditions similar to those we describe.

References

  1. Hypertens Res. 2010 May;33(5):398-410 [PMID: 20379189]
  2. Hypertension. 2015 May;65(5):993-1001 [PMID: 25712718]
  3. Atherosclerosis. 2009 Jan;202(1):296-303 [PMID: 18550063]
  4. Prog Cardiovasc Dis. 2012 Jul-Aug;55(1):14-24 [PMID: 22824106]
  5. J Intern Med. 2016 Jul;280(1):97-113 [PMID: 26749537]
  6. BMC Med Imaging. 2010 Jul 13;10:15 [PMID: 20626888]
  7. Stat Methods Med Res. 2015 Feb;24(1):27-67 [PMID: 24919831]
  8. Diabetes. 2006 Feb;55(2):480-6 [PMID: 16443784]
  9. Nat Methods. 2014 Sep;11(9):895-8 [PMID: 25166868]
  10. BMC Public Health. 2013 Apr 27;13:402 [PMID: 23621920]
  11. Atherosclerosis. 1998 Nov;141(1):133-9 [PMID: 9863546]
  12. BMC Med Imaging. 2014 May 06;14:14 [PMID: 24885423]
  13. BMJ Open. 2017 Dec 14;7(12):e017873 [PMID: 29247091]
  14. PLoS One. 2014 May 09;9(5):e96477 [PMID: 24816787]
  15. Int J Cardiol. 2013 Oct 12;168(5):5026-8 [PMID: 23968713]
  16. Curr Opin Lipidol. 2009 Feb;20(1):57-62 [PMID: 19106708]
  17. J Appl Physiol (1985). 2014 Apr 15;116(8):998-1005 [PMID: 24577061]
  18. Microcirculation. 2014 Aug;21(6):499-505 [PMID: 24589155]
  19. Microcirculation. 2009 Feb;16(2):183-91 [PMID: 19206003]
  20. Opt Express. 2007 Nov 12;15(23):15101-14 [PMID: 19550794]
  21. J Am Soc Nephrol. 2012 Nov;23(11):1900-8 [PMID: 23085635]
  22. Int J Clin Pract. 2010 Feb;64(3):389-403 [PMID: 20456177]
  23. J Clin Epidemiol. 2011 Jan;64(1):96-106 [PMID: 21130355]
  24. Crit Care. 2012 Jun 19;16(3):224 [PMID: 22713365]
  25. Clin J Am Soc Nephrol. 2014 Apr;9(4):698-704 [PMID: 24458084]

MeSH Term

Adult
Cardiovascular Diseases
Erythrocytes
Female
Humans
Longitudinal Studies
Male
Microcirculation
Middle Aged
Mouth Floor
Reproducibility of Results
Video Recording

Word Cloud

Created with Highcharts 10.0.00parametersthreemicrocirculationN=studiesreliabilitysublingualVDPBR95%CI:usedRBCFobtainedsidestreamdarkfieldimagingmeasurementICClimitsagreementintra-ratersetting50conceptmealsmokingReproducibilitysampleBACKGROUND:ChangesmaysurrogateoutcomecardiovasculardiseaseassessedcharacteristicsVascularDensityRedBloodCellFillingPerfusedBoundaryRegionMETHODS:variancecomponentsIntraclassCorrelationCoefficientStandardErrorMeasurementestimatedinter-rater48Subsequentlyproofmeasurespoweranalysisdesignevaluateeffectacutestimuli-iecigarette21RESULTS:poor2measurements2804535127693308-056standarderrorslargerstatisticallysignificantintra-individualinter-individualdifferencesreportedpreviousproofsshowedsizesexcess600subjectsnecessaryreachstatisticalsignificanceobservedeffectsCONCLUSIONS:currentformappearslowlargesizeadvisableuseconditionssimilardescribe

Similar Articles

Cited By