Why low-voltage shock impedance measurements fail to reliably detect insulation breaches in transvenous defibrillation leads.

Charles D Swerdlow, John E Porterfield, Anil G Kottam, Mark W Kroll
Author Information
  1. Charles D Swerdlow: Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California. Electronic address: swerdlow@ucla.edu.
  2. John E Porterfield: Koronis Biomedical Technologies, Maple Grove, Minnesota.
  3. Anil G Kottam: Koronis Biomedical Technologies, Maple Grove, Minnesota.
  4. Mark W Kroll: University of Minnesota, Minneapolis, Minnesota.

Abstract

BACKGROUND: Implantable cardioverter-defibrillators (ICDs) use low-voltage measures of shock impedance (LVSZ) to monitor integrity of leads.
OBJECTIVE: To determine the separation distance between conductors required for LVSZ to detect insulation breaches that produce short circuits during shocks, causing failed defibrillation.
METHODS: We simulated in-pocket insulation breaches between the ICD generator (CAN) and cables to the distal coil of 10 leads from 2 manufacturers. The ICD and lead were placed in an electrolyte bath. Polystyrene sheets were used to control the breach-CAN separation. We determined both the maximum lead-CAN separation for shorts during 800 V shocks and the shock strength at which shorts occurred for a fixed separation. We also calculated breach impedance and measured it using a low-voltage instrument.
RESULTS: The maximum breach-CAN separation for shorting was 350-500 μm for all leads. The minimum shock strength to short varied from 650 to 771 V (24-32 J). LVSZ never triggered a warning, even with no separation between the cable's inner insulation and the CAN. Using low-voltage pulses, breach impedance was measured at approximately 500-1000 Ω.
CONCLUSION: LVSZ is insensitive to insulation breaches that cause life-threatening, shorted shocks. The explanation likely relates to impedance differences between ionic conduction during LVSZ measurements and free-electron conduction in plasma discharges.

Keywords

Grants

  1. R44 AG055056/NIA NIH HHS

MeSH Term

Defibrillators, Implantable
Electric Impedance
Electric Power Supplies
Equipment Design
Equipment Failure Analysis
Humans

Word Cloud

Created with Highcharts 10.0.0separationimpedanceLVSZinsulationlow-voltageshockleadsbreachesshocksImplantabledetectshortdefibrillationICDCANbreach-CANmaximumshortsVstrengthbreachmeasuredconductionmeasurementsBACKGROUND:cardioverter-defibrillatorsICDsusemeasuresmonitorintegrityOBJECTIVE:determinedistanceconductorsrequiredproducecircuitscausingfailedMETHODS:simulatedin-pocketgeneratorcablesdistalcoil102manufacturersleadplacedelectrolytebathPolystyrenesheetsusedcontroldeterminedlead-CAN800occurredfixedalsocalculatedusinginstrumentRESULTS:shorting350-500μmminimumvaried65077124-32Jnevertriggeredwarningevencable'sinnerUsingpulsesapproximately500-1000ΩCONCLUSION:insensitivecauselife-threateningshortedexplanationlikelyrelatesdifferencesionicfree-electronplasmadischargesfailreliablytransvenousDefibrillationImpedancecardioverter-defibrillatorInsulationLeads

Similar Articles

Cited By