3D-printed vaginal cylindrical template with curved needle channels in MRI-guided vaginal cuff brachytherapy.

Zhuang Mao, Hongfu Zhao
Author Information
  1. Zhuang Mao: Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun City, PR China.
  2. Hongfu Zhao: Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun City, PR China.

Abstract

Purpose: To present a new technique for the treatment of vaginal cuff recurrences using 3D-printed vaginal cylindrical template (3D-PVCT) with curved needle channels in brachytherapy.
Material and methods: Two patients with vaginal cuff tumor treated with external beam radiotherapy and MRI-guided interstitial brachytherapy were selected for the present study. Prior to treatment, a vaginal cylinder with the same size as 3D-PVCT was inserted into the vagina. Magnetic resonance scan was performed to identify target volumes and organs at risk (OARs). By registration techniques, the implantation angle was determined, appropriate needle channels were selected, and the depth of each needle position was recorded. During the actual treatment, patients were under general anesthesia, and real-time guidance of trans-rectal ultrasound was applied referring to a pre-determined treatment plan.
Results: For patient No. 1, 9 needles were inserted into the tumor, with 4 curved needle channels and 5 straight channels. For patient No. 2, 7 needles were inserted into the vaginal cuff tumor, with 6 curved needle channels and 1 straight channel. Doses delivered to volumes of targets and OARs for both patients met the EMBRACE II dose constraints. After follow-up of 15 and 18 months, respectively, both patients showed complete response, with no evidence of tumor recurrence. No significant acute or late toxicities were reported.
Conclusions: With careful pre-planning, 3D-PVCT provides good target coverage and sparing of OARs. The results from these two patients indicate that this approach is very promising due to its flexibility and potential widespread application in the future.

Keywords

References

  1. J Contemp Brachytherapy. 2019 Oct;11(5):479-487 [PMID: 31749858]
  2. Radiother Oncol. 2019 Dec;141:123-129 [PMID: 31495516]
  3. Brachytherapy. 2020 Jul - Aug;19(4):438-446 [PMID: 32265118]
  4. Brachytherapy. 2022 Mar-Apr;21(2):193-201 [PMID: 34980568]
  5. Front Oncol. 2022 Jul 04;12:870570 [PMID: 35860574]
  6. Brachytherapy. 2018 May - Jun;17(3):580-586 [PMID: 29426744]
  7. Front Oncol. 2021 Aug 19;11:698773 [PMID: 34490095]
  8. J Cancer Res Ther. 2021 Jul;17(3):688-694 [PMID: 34269300]
  9. Int J Radiat Oncol Biol Phys. 2012 Apr 1;82(5):1880-8 [PMID: 21868174]
  10. Brachytherapy. 2021 Jan-Feb;20(1):85-94 [PMID: 33039332]
  11. J Appl Clin Med Phys. 2020 Aug;21(8):15-26 [PMID: 32459059]
  12. Int J Radiat Oncol Biol Phys. 2023 Mar 1;115(3):654-663 [PMID: 36191741]
  13. Acta Radiol Oncol Radiat Phys Biol. 1978;17(1):33-48 [PMID: 696400]
  14. Radiother Oncol. 2016 Jan;118(1):173-5 [PMID: 26743833]
  15. 3D Print Med. 2019 May 16;5(1):8 [PMID: 31098743]
  16. Radiother Oncol. 2015 Feb;114(2):257-63 [PMID: 25497872]
  17. Brachytherapy. 2023 Jul-Aug;22(4):468-476 [PMID: 37169607]
  18. Radiother Oncol. 2016 Sep;120(3):441-446 [PMID: 27350396]
  19. Int J Radiat Oncol Biol Phys. 2014 Dec 1;90(5):1083-90 [PMID: 25216857]
  20. Lancet Oncol. 2020 Mar;21(3):e157-e167 [PMID: 32135119]
  21. Brachytherapy. 2012 Jan-Feb;11(1):68-75 [PMID: 22265440]
  22. Brachytherapy. 2013 Jul-Aug;12(4):303-10 [PMID: 23491023]
  23. J Radiat Res. 2022 May 18;63(3):402-411 [PMID: 35349700]
  24. J Contemp Brachytherapy. 2022 Aug;14(4):398-402 [PMID: 36199949]
  25. Gynecol Oncol. 2018 Aug;150(2):293-299 [PMID: 29929925]
  26. Int J Radiat Oncol Biol Phys. 2013 Sep 1;87(1):111-9 [PMID: 23849695]
  27. Clin Transl Radiat Oncol. 2018 Jan 11;9:48-60 [PMID: 29594251]
  28. Brachytherapy. 2023 Jul-Aug;22(4):446-460 [PMID: 37024350]
  29. J Contemp Brachytherapy. 2018 Oct;10(5):470-477 [PMID: 30479625]

Word Cloud

Created with Highcharts 10.0.0vaginalneedlechannelspatientstreatmentcuffcurvedbrachytherapytumortemplate3D-PVCTinsertedOARspresent3D-printedcylindricalMRI-guidedinterstitialselectedtargetvolumespatient1needlesstraightPurpose:newtechniquerecurrencesusingMaterialmethods:TwotreatedexternalbeamradiotherapystudyPriorcylindersizevaginaMagneticresonancescanperformedidentifyorgansriskregistrationtechniquesimplantationangledeterminedappropriatedepthpositionrecordedactualgeneralanesthesiareal-timeguidancetrans-rectalultrasoundappliedreferringpre-determinedplanResults:945276channelDosesdeliveredtargetsmetEMBRACEIIdoseconstraintsfollow-up1518monthsrespectivelyshowedcompleteresponseevidencerecurrencesignificantacutelatetoxicitiesreportedConclusions:carefulpre-planningprovidesgoodcoveragesparingresultstwoindicateapproachpromisingdueflexibilitypotentialwidespreadapplicationfuture3Dprintingpre-plan

Similar Articles

Cited By