Inductive plethysmography to control volume-targeted ventilation for leak compensation.

Marie Claire Andrieu, Cedric Quentin, David Orlikowski, Gilbert Desmarais, Daniel Isabey, Bruno Louis, Frédéric Lofaso
Author Information
  1. Marie Claire Andrieu: Saime-ResMed Corporation, Savigny le Temple, France.

Abstract

OBJECTIVE: Volume-targeted ventilation is associated with poor compensation for leaks yet is frequently used in neuromuscular patients, many of whom experience mouth leaks. Our objective was to test the feasibility and efficacy of a prototype ventilator designed to achieve leak compensation by feed-back control of the actual tidal volume based on either the volume blown to the expiratory circuit (VTEVC-compensation) or the tidal volume measured by inductive plethysmography (VTplet-compensation).
DESIGN AND SETTING: Randomized, cross-over, physiological study in a physiological ward of a university teaching hospital.
PARTICIPANTS: Nine normal individuals.
INTERVENTION: Subjects tested volume-targeted assist-control ventilation without compensation, with VTEVC-compensation, and with VTplet-compensation. Tests were done with the mouth closed, with mouth leaks during inspiration, and with mouth leaks during expiration.
MEASUREMENTS AND RESULTS: With inspiratory mouth leaks compared to mouth closed, the delivered volume remained unchanged without leak compensation, increased to 143 +/- 30% with VTEVC-compensation, and increased to 132 +/- 17% with VTplet-compensation; the expired tidal volume decreased to 46 +/- 24% without compensation, 66 +/- 20% with VTEVC-compensation, and 68 +/- 23% with VTplet-compensation. With expiratory mouth leaks, the ventilator-delivered volume and expired tidal volume were unchanged with no compensation and with VTplet-compensation; they increased with VTEVC-compensation.
CONCLUSION: Leak compensation can be achieved during volume-targeted ventilation. VTEVC-compensation and VTplet-compensation were equally effective in compensating for inspiratory leaks, and VTplet-compensation also performed well when expiratory leaks occurred.

References

  1. Chest. 1995 Jun;107(6):1552-8 [PMID: 7781345]
  2. Eur Respir J. 1999 Jan;13(1):157-62 [PMID: 10836341]
  3. Chest. 1988 Feb;93(2):333-5 [PMID: 3276455]
  4. Sleep. 1997 Jul;20(7):561-9 [PMID: 9322272]
  5. Intensive Care Med. 2003 Apr;29(4):596-602 [PMID: 12589533]
  6. Am Rev Respir Dis. 1992 Jan;145(1):114-20 [PMID: 1731573]
  7. Respir Med. 2005 Jan;99(1):52-9 [PMID: 15672849]
  8. Eur Respir J. 2002 Oct;20(4):1029-36 [PMID: 12412700]
  9. Respir Physiol Neurobiol. 2007 May 14;156(2):234-9 [PMID: 17251070]
  10. Eur Respir J. 1999 Dec;14(6):1251-7 [PMID: 10624751]
  11. Arch Bronconeumol. 2004 Nov;40(11):508-17 [PMID: 15530343]
  12. Eur Respir J. 2001 Feb;17(2):259-67 [PMID: 11334129]
  13. Intensive Care Med. 1999 Jul;25(7):662-7 [PMID: 10470568]
  14. Chest. 2005 Nov;128(5):3691-8 [PMID: 16304335]
  15. Chest. 2006 Sep;130(3):815-21 [PMID: 16963680]

MeSH Term

Cross-Over Studies
Equipment Design
Equipment Failure
Humans
Masks
Plethysmography
Pulmonary Gas Exchange
Respiration, Artificial

Word Cloud

Created with Highcharts 10.0.0compensationleaksmouthvolumeVTplet-compensationVTEVC-compensation+/-ventilationtidalleakexpiratoryvolume-targetedwithoutincreasedcontrolplethysmographyANDphysiologicalclosedinspiratoryunchangedexpiredOBJECTIVE:Volume-targetedassociatedpooryetfrequentlyusedneuromuscularpatientsmanyexperienceobjectivetestfeasibilityefficacyprototypeventilatordesignedachievefeed-backactualbasedeitherblowncircuitmeasuredinductiveDESIGNSETTING:Randomizedcross-overstudywarduniversityteachinghospitalPARTICIPANTS:NinenormalindividualsINTERVENTION:Subjectstestedassist-controlTestsdoneinspirationexpirationMEASUREMENTSRESULTS:compareddeliveredremained14330%13217%decreased4624%6620%6823%ventilator-deliveredCONCLUSION:LeakcanachievedequallyeffectivecompensatingalsoperformedwelloccurredInductive

Similar Articles

Cited By