Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients.

Richard Brunner, Reinhard Kitzberger, Wolfgang Miehsler, Harald Herkner, Christian Madl, Ulrike Holzinger
Author Information
  1. Richard Brunner: Division of Gastroenterology and Hepatology - ICU 13H1, Department of Medicine III, Medical University of Vienna, Austria.

Abstract

BACKGROUND: Continuous glucose monitoring has been proposed to optimize glucose control in critically ill patients. To achieve strict glucose regulation, accurate and reliable continuous glucose-monitoring systems are essential.
OBJECTIVE: Evaluation of a subcutaneous continuous glucose-monitoring system for use in critically ill patients.
DESIGN: Pooled-data analysis of two prospective, randomized, controlled trials.
SETTING: An eight-bed medical intensive care unit of a university hospital.
PATIENTS: A total of 174 critically ill patients on intensive insulin therapy.
INTERVENTIONS: Subcutaneous continuous glucose monitoring.
MEASUREMENTS: Two thousand forty-five continuous glucose-monitoring system sensor glucose values were compared with arterial reference blood glucose levels, determined by a blood gas analyzer. Continuous glucose monitoring data were recorded continuously for up to 72 hrs by using a subcutaneous continuous glucose-monitoring sensor. The correlation of both methods and differences between continuous glucose-monitoring systems and reference values were calculated, as well as the conformity of continuous glucose-monitoring values with the International Organization for Standardization criteria (<0.83 mmol/L [15 mg/dL] difference for glucose values ≤ 4.12 mmol/L [≤ 75 mg/dL] and <20% difference for glucose values >4.12 mmol/L [>75 mg/dL]).
RESULTS: The Pearson correlation coefficient was 0.92, showing strong correlation between the two methods. The intraclass correlation coefficient was 0.92, indicating that 92% of the variability is due to subjects and measurement occasions. Mean difference between continuous glucose-monitoring system and reference values was -0.10 mmol/L (confidence interval: -0.13 to -0.07) (-2 mg/dL [confidence interval: -2 to -1]) (continuous glucose-monitoring system minus reference) and absolute difference 0.44 mmol/L (confidence interval: 0.41-0.47) (8 mg/dL [confidence interval: 7-8]). According to the insulin titration error grid analysis, 99.1% of continuous glucose-monitoring system values were in the acceptable treatment zone. No continuous glucose-monitoring system measurements were found in the life-threatening zone, and 92.9% of the continuous glucose-monitoring system glucose values met the International Organization for Standardization criteria.
CONCLUSION: The subcutaneous continuous glucose-monitoring system is reliable for use in critically ill patients and showed glucose values with a strong correlation to arterial reference blood glucose levels, determined by a blood gas analyzer.

MeSH Term

Blood Gas Analysis
Blood Glucose
Critical Care
Critical Illness
Humans
Radiation Monitoring
Reproducibility of Results
Retrospective Studies

Chemicals

Blood Glucose

Word Cloud

Created with Highcharts 10.0.0continuousglucose-monitoringglucosesystemvaluescriticallyillpatientsreferencecorrelationmmol/Lsubcutaneousblooddifference0interval:monitoringmg/dL]92-0ContinuousreliablesystemsuseanalysistwointensiveinsulinsensorarteriallevelsdeterminedgasanalyzermethodsInternationalOrganizationStandardizationcriteria12coefficientstrongconfidence-2mg/dL[confidencezoneBACKGROUND:proposedoptimizecontrolachievestrictregulationaccurateessentialOBJECTIVE:EvaluationDESIGN:Pooled-dataprospectiverandomizedcontrolledtrialsSETTING:eight-bedmedicalcareunituniversityhospitalPATIENTS:total174therapyINTERVENTIONS:SubcutaneousMEASUREMENTS:Twothousandforty-fivecompareddatarecordedcontinuously72hrsusingdifferencescalculatedwellconformity<083[154[≤75<20%>4[>75RESULTS:Pearsonshowingintraclassindicating92%variabilityduesubjectsmeasurementoccasionsMean101307-1]minusabsolute4441-04787-8]Accordingtitrationerrorgrid991%acceptabletreatmentmeasurementsfoundlife-threatening9%metCONCLUSION:showedAccuracyreliability