- M Lettau: Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany. michael lettau@gmx.de
BACKGROUND AND PURPOSE: Contrast-enhanced (CE) 3D magnetic resonance venography (MRV) and CE 3D magnetization prepared rapid acquisition of gradient echo (MP-RAGE) sequences are increasingly commonly used methods for evaluation of the intracranial venous system. Our aim was to compare CE MRV, 2D time-of-flight (TOF) MRV and MP-RAGE sequences at 3 T for visualization of cerebral venous and sinus thrombosis.
MATERIAL AND METHODS: patients with suspected or known cerebral Venous thrombosis were examined prospectively by TOF MRV, CE MRV and MP-RAGE sequences. In 11 consecutive patients (all women; mean age, 42.5 years; age range, 25-70 years) with Venous thrombosis, scores according to overall image quality, and presence or absence (score P) and differentiation (score D) of Venous thrombosis were evaluated.
RESULTS: By all measurements, overall image quality ranged from good to excellent. In 20 of 52 venous structures (38.5%), score P was the same on TOF MRV, CE MRV and MP-RAGE sequences. Venous thrombosis was definitely or almost definitely present or absent with TOF MRV in 20 of 52 (38.5%), with CE MRV in 97 of 99 (97.9%) and with MP-RAGE sequences in 86 of 99 (86.9%) venous structures. In all venous structures with uncertain diagnosis on TOF MRV, thrombosis was definitely or almost definitely present or absent on CE MRV and MP-RAGE sequences. Differentiation of thrombosis was better on CE MRV (score D: 3.33) than on MP-RAGE sequences (score D: 2.78), followed by TOF MRV (score D: 1.32).
CONCLUSION: CE MRV was superior to TOF MRV and MP-RAGE sequences in visualizing cerebral venous and sinus thrombosis.