Adverse Event Causes From 2022 for Four Continuous Glucose Monitors.

Jan S Krouwer
Author Information
  1. Jan S Krouwer: Krouwer Consulting, Sherborn, MA, USA. ORCID

Abstract

BACKGROUND: Adverse events for continuous glucose monitors (CGMs) represent a significant issue for people with diabetes with 281 963 CGM adverse events occurring in 2022. The process to obtain adverse events and the US Food and Drug Administration (FDA) database that contains them are reviewed.
METHODS: Tables were created in SQL Server for four CGM products (Dexcom G6, all versions of Abbott Libre, Medtronic Guardian 3, and Senseonics Eversense) containing either malfunction or injury adverse events sorted by the manufacturer's chosen product code. As the product code is not always clear (or appropriate), the causes of the events were determined from the text description of the adverse event. The resulting causes were listed in decreasing order in tables for each product and event type.
RESULTS: A common effect of several event causes prevented the user from obtaining a result. Inaccuracy was also a frequent complaint. Other causes were specific to that device.
CONCLUSIONS: Creating tables based on manufacturer problem codes for their CGMs, followed by analysis of the adverse event text, facilitates the analysis of event causes. Analyzing adverse event data is the first step in trying to reduce the number of adverse events.

Keywords

References

  1. JAMA. 2020 Jun 16;323(23):2388-2396 [PMID: 32543683]
  2. J Diabetes Sci Technol. 2023 Jan;17(1):252 [PMID: 36214275]
  3. J Diabetes Sci Technol. 2019 Sep;13(5):959-962 [PMID: 30580582]
  4. J Diabetes Sci Technol. 2022 Jan;16(1):228-232 [PMID: 32880188]
  5. Anesthesiology. 1978 Dec;49(6):399-406 [PMID: 727541]

MeSH Term

Humans
Blood Glucose Self-Monitoring
Blood Glucose
United States Food and Drug Administration
United States
Equipment Failure
Databases, Factual
Diabetes Mellitus

Chemicals

Blood Glucose

Word Cloud

Created with Highcharts 10.0.0adverseeventeventscausesCGMproductAdverseCGMs2022malfunctioninjurycodetexttablesanalysisBACKGROUND:continuousglucosemonitorsrepresentsignificantissuepeoplediabetes281963occurringprocessobtainUSFoodDrugAdministrationFDAdatabasecontainsreviewedMETHODS:TablescreatedSQLServerfourproductsDexcomG6versionsAbbottLibreMedtronicGuardian3SenseonicsEversensecontainingeithersortedmanufacturer'schosenalwaysclearappropriatedetermineddescriptionresultinglisteddecreasingordertypeRESULTS:commoneffectseveralpreventeduserobtainingresultInaccuracyalsofrequentcomplaintspecificdeviceCONCLUSIONS:CreatingbasedmanufacturerproblemcodesfollowedfacilitatesAnalyzingdatafirststeptryingreducenumberEventCausesFourContinuousGlucoseMonitorsFMEAMAUDE

Similar Articles

Cited By