Design and Performance Evaluation of SiPM-based High-resolution Dedicated Brain Positron Emission Tomography Scanner: A Simulation Study.

Tahereh Zare, Peyman Sheikhzadeh, Behnoosh Teimourian Fard, Pardis Ghafarian, Mohammad Reza Ay
Author Information
  1. Tahereh Zare: Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Science, Tehran, Iran.
  2. Peyman Sheikhzadeh: Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Science, Tehran, Iran.
  3. Behnoosh Teimourian Fard: Research Center for Molecular and Cellular Imaging (RCMCI), Advanced Medical Technologies and Equipment Institute (AMTEI), Tehran University of Medical Sciences, Tehran, Iran.
  4. Pardis Ghafarian: Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  5. Mohammad Reza Ay: Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Science, Tehran, Iran.

Abstract

Purpose/Aim: The increasing population age highlights the critical need for early brain disease diagnosis, especially in disorders such as dementia. Consequently, a notable focus has been on developing dedicated brain positron emission tomography (PET) scanners, which offer higher resolution and sensitivity than whole-body PET scanners. This study aims to design and performance evaluation of an LYSO-based dedicated brain PET scanner.
Materials and Methods: We developed a dedicated brain PET using Monte Carlo simulation based on cylindrical geometry. Each detector block consisted of a 23 �� 23 array of 2 mm �� 2 mm �� 15 mm LYSO crystals coupled with SiPM. The performance of this scanner was evaluated based on the NEMA NU-2-2018 standard, focusing on analyzing various energy windows and coincidence time windows (CTWs).
Results: The results demonstrated that the noise equivalent count rate (NECR) peaked at each CTW in the 408-680 keV energy window. In addition, increasing the CTWs from 3 ns to 10 ns resulted in a decrease of 9% in sensitivity and an increase of 63% in NECR. Furthermore, the study findings highlight that using a time-of-flight (TOF) resolution of 250 ps can substantially improve image contrast relative to non-TOF reconstruction.
Conclusions: We conclude that employing a broader energy window and a narrower CTW can significantly enhance the scanner's performance regarding sensitivity and NECR. Furthermore, incorporating LYSO pixelated crystals with TOF information will facilitate the generation of high-resolution and high-contrast images.

Keywords

References

  1. Phys Med Biol. 2007 Mar 7;52(5):1505-26 [PMID: 17301468]
  2. Phys Med. 2017 Aug;40:59-65 [PMID: 28716541]
  3. IEEE Trans Radiat Plasma Med Sci. 2019 May;3(3):334-342 [PMID: 31453423]
  4. Phys Med Biol. 2018 Sep 10;63(18):185005 [PMID: 30113313]
  5. Med Phys. 2019 Nov;46(11):4816-4825 [PMID: 31448421]
  6. Phys Med. 2024 Jul;123:103395 [PMID: 38843650]
  7. J Nucl Med. 2014 May 12;55(Supplement 2):2S-10S [PMID: 24819419]
  8. Med Phys. 2023 Nov;50(11):6815-6827 [PMID: 37665768]
  9. EJNMMI Phys. 2016 Dec;3(1):3 [PMID: 26879863]
  10. J Nucl Med. 2015 Jan;56(1):98-105 [PMID: 25525181]
  11. Phys Med Biol. 2012 Feb 21;57(4):867-83 [PMID: 22290410]
  12. IEEE Trans Nucl Sci. 2014 Jun 1;61(3):1164-1173 [PMID: 25774063]
  13. Lancet Neurol. 2008 Sep;7(9):812-26 [PMID: 18667359]
  14. J Nucl Med. 2016 Apr;57(4):646-52 [PMID: 26697961]
  15. Alzheimers Dement. 2013 Jan;9(1):63-75.e2 [PMID: 23305823]
  16. Phys Med Biol. 2019 Nov 26;64(23):235004 [PMID: 31618708]
  17. Comput Med Imaging Graph. 2014 Jul;38(5):358-68 [PMID: 24766905]
  18. Phys Med Biol. 2019 Oct 31;64(21):215013 [PMID: 31530762]
  19. J Instrum. 2014 May 12;9(5): [PMID: 24883078]

Word Cloud

Created with Highcharts 10.0.0brainPETsensitivitydedicatedperformance��mmenergyNECRincreasingpositronemissiontomographyscannersresolutionstudyscannerusingMonteCarlosimulationbased232LYSOcrystalswindowstimeCTWsnoiseequivalentcountrateCTWwindownsFurthermoreTOFcanDedicatedPurpose/Aim:populationagehighlightscriticalneedearlydiseasediagnosisespeciallydisordersdementiaConsequentlynotablefocusdevelopingofferhigherwhole-bodyaimsdesignevaluationLYSO-basedMaterialsMethods:developedcylindricalgeometrydetectorblockconsistedarray15coupledSiPMevaluatedNEMANU-2-2018standardfocusinganalyzingvariouscoincidenceResults:resultsdemonstratedpeaked408-680keVaddition310resulteddecrease9%increase63%findingshighlighttime-of-flight250pssubstantiallyimproveimagecontrastrelativenon-TOFreconstructionConclusions:concludeemployingbroadernarrowersignificantlyenhancescanner'sregardingincorporatingpixelatedinformationwillfacilitategenerationhigh-resolutionhigh-contrastimagesDesignPerformanceEvaluationSiPM-basedHigh-resolutionBrainPositronEmissionTomographyScanner:SimulationStudyflight

Similar Articles

Cited By