Change in irrigation flow through a flexible ureteroscope with various devices in the working channel: Comparison between an automatic irrigation pump and gravity-based irrigation.
Takaaki Inoue, Fukashi Yamamichi, Shinsuke Okada, Shuzo Hamamoto, Masato Fujisawa, SMART Study Group
Author Information
Takaaki Inoue: Department of Urology and Surgery, Kobe University, Hyogo, Japan.
Fukashi Yamamichi: Department of Urology, Hara Genitourinary Hospital, Hyogo, Japan.
Shinsuke Okada: Department of Urology, Gyotoku General Hospital, Chiba, Japan.
Shuzo Hamamoto: Department of Nephro-Urology, Nagoya City University Graduate School of Medical Sciences, Aichi, Japan.
Masato Fujisawa: Department of Urology and Surgery, Kobe University, Hyogo, Japan.
OBJECTIVES: To evaluate the change in the irrigation flow with various instruments in the working channel of a flexible ureteroscope by two automatic irrigation pumps and gravity-based irrigation in an ex vivo setting. METHODS: We used two automatic irrigation pumps: the Endoflow II and the UROMAT Endoscopic Automatic System for Irrigation and gravity-based irrigation. A flexible ureteroscope was connected to an irrigation tube with a working channel. The other side of the irrigation tube was attached to each automatic irrigation pump, which was connected with a 2-L saline bag or to a 2-L saline bag directly in case of gravity pressure. The flow volume from the working channel was measured three times for 30 s at various irrigation pressure settings, both when the working channel was unoccupied and occupied with various instruments. RESULTS: The irrigation flow steadily increased as the irrigation pressure in the automatic irrigation pumps increased and the saline position in gravity became higher (P < 0.05). However, the flow decreased as the size of the instrument in the working channel increased (P < 0.05). The efficiency of irrigation flow in gravity-based irrigation under the same pressure is significantly lower than one of two automatic irrigation pumps (P < 0.05). However, there was no significant difference in the efficiency of the irrigation flow between the Endoflow II and UROMAT Endoscopic Automatic System for Irrigation. The irrigation pressure setting needed to change to maintain adequate irrigation flow when using various working tools. CONCLUSIONS: The efficiency of irrigation flow in gravity-based irrigation is significantly lower than one of two automatic irrigation pumps. The irrigation flow decreases as the size of the instrument in the working channel increases.
Vlot J, Wijnen R, Stolker RJ, Bax KN. Optimizing working space in laparoscopy: CT measurement of the effect of pre-stretching of the abdominal wall in a porcine model. Surg. Endosc. 2014; 28: 841-6.
Lama DJ, Owyong M, Parkhomenko E, Patel RM, Landman J, Clayman RV. Fluid dynamic analysis of hand-pump infuser and UROMAT endoscopic automatic system for irrigation through a flexible ureteroscope. J. Endourol. 2018; 32: 431-6.
Thomsen HS, Talner LB, Higgins CB. Intrarenal backflow during retrograde pyelography with graded intrapelvic pressure: a radiologic study. Invest. Radiol. 1982; 17: 593-603.
Thomsen HS. Pyelorenal backflow. Clinical and experimental investigations. Radiologic, nuclear, medical and pathoanatomic studies. Dan. Med. Bull. 1984; 31: 438-57.
Blew BD, Dagnone AJ, Pace KT, Honey RJ. Comparison of peditrol irrigation device and common methods of irrigation. J. Endourol. 2005; 19: 562-5.
Abdelshehid C, Ahlering MT, Chou D et al. Comparison of flexible ureteroscopes: deflection, irrigant flow and optical characteristics. J. Urol. 2005; 173: 2017-21.
Bach T, Geavlete B, Herrmann TR, Gross AJ. Working tools in flexible ureterorenoscopy-influence on flow and deflection: What does matter? J. Endourol. 2008; 22: 1639-43.
Lechevallier E, Luciani M, Nahon O, Lay F, Coulange C. Transurethral ureterorenolithotripsy using new automated irrigation/suction system controlling pressure and flow compared with standard irrigation: a randomized pilot study. J. Endourol. 2003; 17: 97-101.
Nagele U, Horstmann M, Hennenlotter J et al. Size does matter: 1.5 Fr. stone baskets almost double irrigation flow during flexible ureteroscopy compared to 1.9 Fr. Stone baskets. Urol. Res. 2006; 34: 389-92.